| Specification
| PM10
| PM2.5
| PM10-2.5
|
|---|
| Class I
| Class II
| Class III
| Class II
| Class III
|
|---|
| Acceptable concentration range (Rj), µg/m
3 | 5-300 | 3-200 | 3-200 | 3-200 | 3-200 | 3-200.
|
| Minimum number of test sites | 2 | 1 | 2 | 4 | 2 | 4.
|
| Minimum number of candidate method samplers or analyzers per site | 3 | 3 | 3
1 | 3
1 | 3
1 | 3.
1
|
| Number of reference method samplers per site | 3 | 3 | 3
1 | 3
1 | 3
1 | 3.
1
|
| Minimum number of acceptable sample sets per site for PM10 methods:
| | | | | | |
| Rj < 20 µg/m
3 | 3 | | | | | |
| Rj > 20 µg/m
3 | 3 | | | | | |
| Total | 10 | | | | | |
| Minimum number of acceptable sample sets per site for PM2.5 and PM10-2.5 candidate equivalent methods:
| | | | | | |
| Rj < 15 µg/m
3 for 24-hr or Rj < 8 µg/m
3 for 48-hr samples. | | 3 | 3 | 3 | 3 | 3.
|
| Rj > 15 µg/m
3 for 24-hr or Rj > 8 µg/m
3 for 48-hr samples | | 3 | 3 | 3 | 3 | 3.
|
| Each season | | 10 | 23 | 23 | 23 | 23.
|
| Total, each site | | 10 | 23 | 23 (46 for two-season sites) | 23 | 23 (46 for two-season sites).
|
| Precision of replicate reference method measurements, PRj or RPRj, respectively; RP for Class II or III PM2.5 or PM10-2.5, maximum | 5 µg/m
3 or 7%. | 2 µg/m
3 or 5%. | 10%
2 | 10%
2 | 10%
2 | 10%.
2
|
| Precision of PM2.5 or PM10-2.5 candidate method, CP, each site | | | 10%
2 | 15%
2 | 15%
2 | 15%.
2
|
| Slope of regression relationship | 1 ±0.10 | 1 ±0.05 | 1 ±0.10 | 1 ±0.10 | 1 ±0.10 | 1 ±0.12.
|
| Intercept of regression relationship, µg/m
3 | 0 ±5 | 0 ±1 | Between: 13.55—(15.05 × slope), but not less than—1.5; and 16.56—(15.05 × slope), but not more than +1.5 | Between: 15.05—(17.32 × slope), but not less than—2.0; and 15.05—(13.20 × slope), but not more than +2.0 | Between: 62.05—(70.5 × slope), but not less than—3.5; and 78.95—(70.5 × slope), but not more than +3.5 | Between: 70.50—(82.93 × slope), but not less than—7.0; and 70.50—(61.16 × slope), but not more than +7.0.
|
| Correlation of reference method and candidate method measurements | ≥ 0.97 | ≥ 0.97 | ≥ 0.93—for CCV ≤ 0.4;
≥ 0.85 + 0.2 × CCV—for 0.4 ≤ CCV ≤ 0.5;
≥ 0.95—for CCV ≥ 0.5
|