§ 213.63 - Equipment design, size, and location.

Equipment used in the manufacture, processing, packing, or holding of a medical gas shall be of appropriate design and adequate size, and be suitably located to facilitate operations for its intended use and any necessary cleaning and maintenance.

§ 213.65 - Equipment construction.

(a) Equipment shall be constructed so that surfaces that contact components, in-process materials, or medical gases shall not be reactive, additive, or absorptive so as to alter the safety, identity, strength, quality, or purity of the medical gas beyond the official or other established requirements.

(b) Any substances required for operation, such as lubricants or coolants, shall not come into contact with components, containers, closures, in-process materials, or medical gases so as to alter the safety, identity, strength, quality, or purity of the medical gas beyond the official or other established requirements.

§ 213.67 - Equipment maintenance and cleaning.

(a) Written procedures shall be established, maintained, and followed for adequate cleaning and maintenance of equipment used in the manufacture, processing, packing, or holding of medical gases. These procedures shall include, but are not necessarily limited to, the following:

(1) Assignment of responsibility for cleaning and maintaining equipment;

(2) Maintenance and cleaning schedules, including, where appropriate, sanitizing schedules;

(3) A description in sufficient detail of the methods, equipment, and materials used in cleaning and maintenance operations, and the methods of disassembling and reassembling equipment as necessary to assure proper cleaning and maintenance;

(4) Removal or obliteration of previous batch identification;

(5) Protection of clean equipment from contamination prior to use; and

(6) Inspection of equipment for cleanliness immediately before use.

(b) The procedures described in paragraph (a) of this section shall not alter the safety, identity, strength, quality, or purity of the medical gas beyond the established requirements.

(c) Records shall be kept of cleaning, maintenance, and inspection as specified in §§ 213.180 and 213.182.

§ 213.68 - Automatic, mechanical, and electronic equipment.

(a) Automatic, mechanical, and electronic equipment used in the manufacture, processing, packing, and holding of medical gases shall be routinely calibrated, inspected, and checked according to a written program designed to ensure proper performance. Written procedures and records of calibration, inspections, and checks shall be maintained.

(b) Computerized systems that record, store, or use data shall be appropriately validated.

(c) A backup file of data entered into the computer system shall be maintained except where certain data, such as calculations performed in connection with laboratory analysis, are eliminated by computerization or other automated processes.

(d) Appropriate change control shall be used whenever modifications are made to computer systems to assure that any changes do not adversely affect data integrity or product quality. Records of such modifications shall be maintained.

source: 89 FR 51770, June 18, 2024, unless otherwise noted.
cite as: 21 CFR 213.68